Since Cast 001 & 002 O2s were very different than 003-066, I compared the .con
O2 calibration coefficients. They are the same so the sensor took two casts to
stabilize.

9507001.con O2

                      oxygen, current
-------------------------------------
 Serial Number =  13204
 Calibration Date =  01-Jun-95
 M =  4.1188000E-07
 B =  -8.2377000E-10
 Soc =  2.7213
 Boc =  -0.0235
 Tcor =  -3.300000E-02
 Pcor =  1.500000E-04
 Tau =  2.00000000E+00
 Wt =  0.670
 
                     oxygen, temperature
---------------------------------------
 K =  4.0431
 C =  15.5206
 
 
 9507002.con
                       oxygen, current
-------------------------------------
 Serial Number =  13204
 Calibration Date =  01-Jun-95
 M =  4.1188000E-07
 B =  -8.2377000E-10
 Soc =  2.7213
 Boc =  -0.0235
 Tcor =  -3.300000E-02
 Pcor =  1.500000E-04
 Tau =  2.00000000E+00
 Wt =  0.670
 
                     oxygen, temperature
---------------------------------------
 K =  4.0431
 C =  15.5206
 
 9507003.con
 
                       oxygen, current
----------------------------------------
 Serial Number =  13204
 Calibration Date =  01-Jun-95
 M =  4.1188000E-07
 B =  -8.2377000E-10
 Soc =  2.7213
 Boc =  -0.0235
 Tcor =  -3.300000E-02
 Pcor =  1.500000E-04
 Tau =  2.00000000E+00
 Wt =  0.670
                     oxygen, temperature
---------------------------------------
 K =  4.0431
 C =  15.5206
 
 