CalCOFI 9702JD CTD Data (Re)Processing Notes

This cruise with single T, C, & O2 sensors looks clean with no obvious sensor problems.
The fluorometer sensitivity was changed after cast 008, probaby from 1x to 10x. These
data were plotted separately - 001-008 vs chla, 009-070 vs chla.

Raw CTD files were converted to cnvs using DOS datcnv.exe under WindowsXP/DOS. 
I tried to use a rebuilt .xmlcon file using coefficients entered from the .con
file but the O2 data were unusually high. So I did the .cnv files in Seasoft for
DOS and it worked fine. All other modules were applied using Windows SBE Data Processing
modules.

There were four additional stations 067, 068, 069, & 070. These appear to be northern
NOAA-requested stations with bottle samples. So the IEH had samples to correct the 
sensor data.

JRW  12/23/2021