CalCOFI 9709NH had no real sensor issues that were detected. The Seabird 911
sensor configuration only had primary sensors. This was before dual T, C, &
O2 sensors were deployed. There are secondary sensor data or plots.
Configuration: one Temperature sensor, one conductivity sensor, one Beckman/YSI
oxygen sensor, fluorometer, transmissometer, PAR, & surface Par.
V0: transmissometer; V1: fluorometer; V2: Altimeter; V3: Biospherical PAR; 
V4/5: Beckman/YSI oxygen sensor; V9 may have been PAR.

Note: this cruise required the DOS-version of the SBE data processing software
to convert the .dat to .cnv. 
This required using DOS Seasoft under a virtual Windows XP in Windows 7. 
Then the data were processed using Windows 7.23 version of the SBE Data Processing 
Suite. Derive required creating a .xmlcon from the .con file. 
Since there appeared to be no sensor changes, only one .xmlcon was used for all the stations.

Corrections: Cast 001 - 93.26.7 line.station number was changed to 93.3 26.7
Cast 050 - 83.40.6 line.station number was changed to 83.3 40.6

No data were flagged as questionable or bad.

There were some special cast labeled "600s". Refer to the associated .hdr for
specific info.

JRW 10/07/2021