CalCOFI 0610RR CTD Data Processing Notes (reprocessed 01/24/2019, adding Quality Codes & uM/Kg Oxygens)
General Notes: We started data acquisition on deck BEFORE deploying and 10m soak. It is necessary to
  run Loopedit to remove the noisy data from 0-10m "2-minute" soak before returning back to surface.
  Some shallow casts (002, 003, 030, 032, 053, 054, 057, 058, 075) were not loopedited since it 
  eliminated too much data. These casts were manually scanned for the appropriate number of scans to
  skip then binavg.
  Cast 002 - had no 10m soak so data acquisition was not started right away.
  Cast 003 - 19359 scans were skipped when binavg, no loopedit
  Cast 030 - 11960 scans were skipped when binavg, no loopedit
  Cast 032 - 11235 scans were skipped when binavg, no loopedit
  Cast 053 - 15884 scans were skipped when binavg, no loopedit
  Cast 054 - 36660 scans were skipped when binavg, no loopedit
  Cast 057 - 7410 scans were skipped when binavg, no loopedit
  Cast 058 - 12270 scans were skipped when binavg, no loopedit
  Cast 075 - 14425 scans were skipped when binavg, no loopedit
  
CTD Setup: 

  Transmissometer important note: 
    V0 - Tr (CST490DR), new M & B were calculated from the min/max voltages in air from the 8 deck tests.
    Min voltage = 0.0611, Max voltage changed but highest (Deck001) was 4.7534
    (Deck013 = 4.7521; 021 = 4.7399; 043 = 4.7143; 048 = 4.6581; 055 = 4.703; 067 = 4.6923; 071 = 4.6923)
    M & B in the individual .con files were from last cruise so a new .con file (0610UseForAllTr.xmlcon) was 
    used on all stations since no sensors were changed during the cruise.
  
  V1 - Seapoint Fluorometer set at 10x
  V2 - ISUS Nitrate Sensor
  V3 - open
  V4 - SBE43 Oxygen Sensor (only one deployed pre-2009)
  V5 - open
  V6 - Benthos/Datasonic Altimeter
  V7 - PAR
  V15 - SPAR
  Sensors: T1 - SN1324; T2 - SN6049; C1 - SN2206; C2 - SN722; O2 - SN680; PAR - 4544; SPAR - SN6369
  
Casts 034 - 039 hdr & dat had wrong station number 86.8 should be 86.7, corrected in both locations, archived, & posted
Cast 036 - Santa Monica Basin station, plotted to 800m
Cast 537 - student cast to 1025m, no bottles, plotted cruise_corr O2, estChl, & estNO3

Quality Coded Casts:
Cast 008U - ISUS data from 270-320 -99'd in the asc file so the regression could do an estimated ISUS NO3
Cast 011UD - all ISUS data QC'd "9" (bad)
Cast 027U - ISU data from 0-125m QC'd "9"
Cast 028UD - all ISUS data "9"
Cast 039D - ISUS data 0-13m "9"
Cast 043D - ISUS data 0-70m "8"
Cast 043U - all ISUS data "9"
Cast 044UD - all ISUS data "9"
Cast 048D - ISUS data 139-268m "9"
Cast 048U - all ISUS data "9"
Cast 066UD - all ISUS data "9"
Cast 067UD - all ISUS data "9"