CalCOFI 1307NH CTD Data Processing Notes
CalCOFI 1307NH: 74 stations occupied, 66 standard & 8 SCCOOS (80.0 50.5 dropped due to weather); station order was impacted by Naval Operations. 
CTD configuration was standard: Seabird 911+ with dual T, C, O2, & pumps; Wetlabs C-Star 25cm transmissometer; Biospherical QSP200L PAR; Datasonics/Benthos Altimeter; NEW ADDITION is a brand new WET Labs ECO-AFL/FL, replacing our Seapoint Fluorometer. Sensitivity setting was 0-50mg/m^3; Seabird pH; Satlantic ISUS v2 & battery. Please refer to the xmlcon files or cruise  prospectus for additional info.
This cruise had many non-confirmation problems - bottles being closed but no confirmation returning. It was later discovered faulty conductive wire(s) were to blame after the questionable wires were excluded on only a single conductor and shield ground used.

Cast 001 93.3 26.7: CTD started down but archiving was not initiated so it was brought back to surface, data acquistion started then lowered

Cast 002 93.4 26.4: No trip at #3, tripped 2 at 5m. Omit #4, re-numbered #5 to surface bottle.

Cast 007 93 40- Bottle 11 didnt confirm, fired again and the next bottle to be fired advanced by 2 numbers. 2 bottles were tripped at 100m.

Cast 008 93.3 45.0- 270m #5, non confirm_ next bottle jumped from 5 to 7.

Cast 009 93.3 50.0: #12 85m no confirmation - updated to #14 next depth

Cast 012 93 70- Really high NO3, Non confirm 1 and 2, finally caught up at 3

Cast 013 93.3 80.0- Non confirms at #s 6, 11, 15, 16,17, 18 went from 16-18should be 19? Think #16 was a non closure 3 bottles still open at surface, sent back down to trip 48 m for prodo with # 22.

Cast 014 93 90- 2 non confirms at 380m non confirm at 270m and non confirms at 50m and 40m, Duplicates at 380 m and no 40m bottle

Cast 021 - 3500m deep cast "Scan length error: expected = 88, actual = 15" message seen during the upcast on-screen at 657m. No CTD sensor data collected between 1758m-1267m data on upcast - signal failure that restarted at 1266m to surface. Another glitch: oxygen upcast profile from ~1800-1300m looks wrong

Cast 023 90.0 70.0: Eventlog GPS wrong, computer serving GPS frozen

Cast 032 86.7 110.0- 4 bottles tripped at 29 m extras are 19 20 21 22 brought on deck reset 19-22; put back into water after resetting

Cast 039 83.3 55.0: bottle #21 mistripped, no closure

Cast 044 81.9 46.9- Popped 2 bottles at 20m, pulled CTD on deck emptied bottle 23 and sent back in to sample 10m depth. No ph data.

Cast 046 80 55- 15 m DIC Sample.

Cast 047 80 60-No ISUS Data.

Cast 050 80 90- Transmissometer spikey near terminal depth at 400-515m.

Cast 054 77 80- Inversion at 50m

Cast 062 87.45: ISUS sent in unplugged, returned to deck immediately but did not work

Cast 063 87.40: ISUS still not working even though the battery was replaced; went from 765 to 700 then back to 750 on upcast to get 15m-off-bottom bottle

Cast 064 87.35: still no ISUS data; down-welled O2 feature at 250m

Cast 065 85.4 35.8: No ISUS data

Cast 066 87.33: No ISUS data

Cast 067 86.8 32.5: ISUS data finally, main cable questionable

Cast 062-69: No reliable ISUS data.

Cast 073 90 37- Bottle 20 did not close, sent CTD back down and used bottle 24 to sample 20m 