1604 CTD-Bottle Issues

Cast 001 - 007 original xmlcon files had incorrect secondary oxygen coefficients.
Also, it was discovered the 2015 service of the CTD pressure sensor resulting in
new pressure coefficients had not been applied. So on cast 026, the pressure 
coefficients were imported. All previous xmlcon files (001-025) were replaced by 
cast 026 xmlcon, relabeled to match the cast number. The pressure data calculated
using the old coefficients were compared to pressure data using the new coefficients
and the difference was barely preceptible ie calibration had changed very little.
Still we prefer to use the latest coefficients so these were applied to earlier
casts. It did not change the bottle stop depths or deck pressure significantly.

Cast 086, 087, 090 mrk files had many extra surface markers which caused the 
  sta.csvs to have issues. These sta.csvs were manually corrected to the right
  number of bottles and repopulated with CTD btl & asc data, plus salts, oxys,
  chls (no nuts here).
  
Bad samples or mistrips:

Salt run for stas 036-039 had a bad sample count (29, should be 28), salts rerun
  and sta.csvs repopulated.
Cast 010 bottle #6 230m questionable NO3 (duplicate of #5?)
Cast 015 CTD upcast & downcast between 50-100m different - looks real
Cast 049 bottle #16 67m bad bottle salt, flagged
Cast 052 bottle #10 120m bad bottle salt, flagged
Cast 072 bottle #10 120m mistrip - all samples bad, flagged
Cast 073 bottle #17 30m bad bottle salt, flagged

CTD Console Notes:
Cast 001 - Messing with the rigging of the conductive wire near bottle #4 precast caused two issues: bottle #4 bottom cap was not open (seen closed during deployment) so bottle #5 was closed at 28m. ISUS battery was not plugged in so no ISUS data.
Cast 002 - RV Shimada Captain did not want to go shallower than 30m so 20m SCCOOS station was at 30m
Cast 008 - noticable upcast O2 hysteresis
Cast 017 - deep chl max ~130m; secchi depth 26m; deepest prodo bottle depth 91m
Cast 018 - unusual salinity stratification ~100-150m; deep chl max again ~127m
Cast 021 - early prodo station so nets first then ran back to sta; moderate seas with big swell, sunny
Cast 025 - Calm, overcast; Mola Molas and Velellas (Mola eating Velella)
Cast 031 - nets-first early prodo again, ~0.58nm off station which we thought was fine but much deeper ~200m instead of ~70m; altimeter reading not stable till ~190m (slope?)
Cast 032 - SCCOOS 88.5 30.1 at 20m
Cast 038 - ISUS battery not connected, no ISUS data
Cast 039 - new ISUS battery installed
Cast 040 - Santa Monica Basin prodo station, only two bottles available for sampling below 515m; light rain
Cast 044 - bottles 14 & 15 had #14 lanyard stuck in the top of the bottles; bottom lanyard clip from #14 inside #15 so probably not rigged
Cast 045 - computer network issue pre-cast, short delay; new winch operator training - CTD held at ~4m on downcast for ~2mins during downcast startup
Cast 049 - unusual salinity & temperature features between ~30-150m; prodo bottles hit chl max; bluewater & sunny
Cast 053 - calculated 6m bottle spacing around 100m changed so an extra bottle at the chl max for NCOG could be sampled
Cast 054 - "interesting" salt & O2 features between ~50-250m
Cast 055 - fluorometer is spikey
Cast 056 - fluorometer is spikey at depth, most likely noise, not real
Cast 058 - Santa Barbara Basin prodo cast, only two bottles available to sample below 515m; fluorometer still spikey after cable servicing; deep NO3 and O2 higher than 500m, basin must have flushed recently
Cast 064 - rough morning
Cast 065 - rough, sunny, windy prodo station
Cast 069 - moderate seas, overcast, big rolls, prodo sta
Cast 070 - ADK training on CTD console
Cast 072 - sta 80.60 pretty rough conditions but doable - sta 80.80 & 80.70 cancelled due to conditions being too rough
Cast 073 - last standard 75-sta pattern station (73 of 75 stations completed)
Cast 074 - on line 73.3, after finishing line 80 near coast; switched to 12-bottle rosette configuration, table-driven; DMW still onboard, most SIO-CalCOFI disembarked
Cast 077 - pH & PAR caps left on - no data; bottle #2 mistripped, bottom cap ajar
Cast 078 - bottle #10 didn't seal properly, leaking
Cast 081 - bottle #10 replaced pre-cast, unreliable
Cast 083 - soapy grey water flowing off deck during the cast
Cast 086 - hdr mislabeled 76.7 50.0 instead of 66.7 50.0; fixed
Cast 087 - hdr mislabeled 76.7 55.0 instead of 66.7 55.0; fixed
Cast 088 - hdr mislabeled 76.7 60.0 instead of 66.7 60.0; fixed
Cast 090 - ISUS battery not plugged in, no ISUS data; battery terminal fried - replaced battery postcast with new one
Cast 099 - wire angle changing a lot, current?, CTD bottle target depths adjusted repeatedly